LEGALDIGMLEGALDIGM
    ← Code of Federal Regulations
    Beta
    10 CFR § 35.457

    10 CFR § 35.457 — § 35.457 Therapy-related computer systems.

    Beta. Legal information, not legal advice. LEGALDIGM is not a law firm. Check the official source for the current text.

    Analyze this law with AI
    § 35.457 Therapy-related computer systems. The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of: (a) The source-specific input parameters required by the dose calculation algorithm; (b) The accuracy of dose, dwell time, and treatment time calculations at representative points; (c) The accuracy of isodose plots and graphic displays; and (d) The accuracy of the software used to determine sealed source positions from radiographic images.

    Related guides, FAQ and contract templates